REPLACING THE GRINDING WHEEL
1. To replace the grinding wheel follow the steps for removing the grinding wheel (see
page26).Afterthegrindingwheelisremoved,youneedtoremovethethreeat-head
screws holding the drive plate adapter onto the diamond grinding wheel.
Note: When you purchase a replacement Grinding Wheel (Part # DSQ0500), it does not come
with the Drive Plate Adapter (Part # 801381) or Flat Head Screws (Part # 801382).
These parts need to be ordered separately or re-used if still in good condition.
2. CleantheinsideoftheDrivePlateAdapterandcheckforwear.Ifthesquarearboris
worn,itwillpreventthegrindingwheelfromoatingontheangenut,andshouldbe
replaced.
3. Place the Drive Plate Adapter onto the new Diamond Grinding Wheel and align the
holes. A small drop of Loctite thread lock should be applied to each screw to keep it
from loosening. Tighten each screw until seated, then go back around and re-tighten
each screw to secure them evenly.
4. The new Diamond Grinding Wheel can now be reinstalled on the tool.
Grinding wheel, adapter and screws
Figure-27
Drive Plate
Adapter
Grinding Wheel
(26mm Arbor)
Flat Head Screws
(M4x6)