Alpha 2000 Uživatelský manuál Strana 13

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 24
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 12
Thermo Scientific / Alpha Process Products / 2000 Series Controllers / www.thermo.com/process 13
Specification Information
Conductivity Controller/Transmitter
Alpha COND 2000W
(Wall mount)
Alpha COND 2000P
(Panel mount)
Order Code TSCONCTP2000W TSCONCTP2000P
Part No. 01X275376 01X275377
Conductivity:
Range:
... to 2.000 µS/cm ; ... to 20.00 µS/cm ;
... to 200.0 µS/cm ; ... to 2000 µS/cm ;
... to 20.00 mS/cm ; ... to 200.0 mS/cm ;
... to 1000 mS/cm
Resolution:
0.001 µS/cm ; 0.01 µS/cm ;
0.1 µS/cm ; 1 µS/cm ;
0.01 mS/cm ; 0.1 mS/cm ;
1 mS/cm
Accuracy: ±1 % of full scale reading (±2 % >500 mS/cm)
Temperature:
Range: -10.0 to 125.0 °C (14.0 to 257.0 °F)
Resolution: 0.1 ºC / 0.1 ºF
Accuracy: ±0.5 ºC / ±1.0 ºF
Sensor: Pt100/Pt1000 (jumper selectable); 2 or 3 wire
Compensation: Auto/manual (normalized at 25 °C)
Set-point & controller functions:
Set point 1 (SP1) / set point 2 (SP2):
... to 2.000 µS/cm or ... to 20.00 µS/cm or
... to 200.0 µS/cm or ... to 2000 µS/cm or
... to 20.00 mS/cm or ... to 200.0 mS/cm or
... to 1000 mS/cm
Switching Conductivity hysteresis: 0 to 10 % of full scale
Function (switchable): P/PI control (pulse length/pulse frequency/proportional integral); limit control; off
Adjustable period with pulse
length controller:
0.5 to 20 sec
Adjustable period with pulse
frequency controller:
60 to 120 pulse/min
Integral action time (IAT): 0 to 999.9 min
Pickup/dropout delay: 0 to 2000 sec
Wash cycle: 0.1 to 200.0 hr
Wash duration: 1 to 2000 sec
Contact outputs: 1 SPDT, 3 SPST relays
Switching voltage/current/power:
1
/
8
HP: at 125 VAC / max. 0.74 A / max. 93 VA
1
/
8
HP: at 250 VAC / max. 0.37 A / max. 93 VA
Alarm functions:
Function (switchable): Steady or fleet (pulse)
Pickup delay: 0 to 2000 sec
Switching voltage/current/power:
1
/
8
HP: at 125 VAC / max. 0.74 A / max. 93 VA
1
/
8
HP: at 250 VAC / max. 0.37 A / max. 93 VA
Electrical data & connections:
Transmitter function: Two 0/4 to 20 mA scalable outputs for Conductivity and temperature, galvanically isolated
CU 22 function: 22 mA current output
Isolated output voltage: 12 V ±0.5 V (max. 50 mA)
Hold function switch: To freeze output current and deactivate control relays
Load: Max. 600 Ω
Conductivity input: 2-cell/4-cell terminal
Connection terminal:
5-pin, 8-pin, 9-pin & 13-pin terminal,
detachable blocks
5-pin, 9-pin & 19-pin terminal,
detachable blocks
Display:
LCD: UV coat, backlit 14 segments display with symbols for status information
Backlight: On/off selectable with four level of brightness control
Power supply:
Input: 80 to 250 VAC/DC ; 50/60 Hz ; 10 VA
Main fuse: 315 mA, anti-surge, 250 V
Pollution degree: 2
Transient overvoltage category: II
EMC specifications:
Emitted interference: According to EN 61326
Immunity to interference: According to EN 61326
Environmental conditions:
Operating temperature range: 0 to 40 °C
Max. relative humidity: 80 % up to 31 °C decreasing linearly to 50 % at 40 °C
Mechanical specifications:
Dimensions (WxHxD): 144 x 144 x 110 mm 96 x 96 x 175 mm
Weight: 745 g (unit) / 1100 g (packed) 550 g (unit) / 950 g (packed)
Ingress protection: IP66 (NEMA 4X) IP54 (front panel)