Figure-7 Bulb View of the Alpha-1000
Lamp Adjustment
The position of your lamp has been factory adjusted to its optimal setting. Follow
these steps below for adjusting the lamp position. These steps should NOT be
needed! Unfortunately, one should COMPLETELY close the light chamber
between adjustment settings attempts. The bulb is driven by high voltage. Plus,
the light output is blinding and will cause eye damage if directly viewed with the
unprotected eye.
1. Loosen the screw on the lamp panel to reveal the light bulb within the lamp
housing.
2. Make certain that the light bulb is fully inserted into its holder. The bulb socket
has been permanently mounted in the optimal position by the factory. So the
rest of steps should not be needed!
3. The lamp socket is mounted on an “L”-shaped bracket. This bracket is held in
place by 2 screws. Slightly loosen the 2 screws such that you may move the
bracket a little. You should then re-tighten the screws, completely. Close the
lamp chamber and then power up. Set your spectrophotometer to %T. Continue
to adjust the bracket position so that you reach the optimum transmittance
reading.
Tip: Don’t loosen the screws too much as you adjust the bracket. Very small
movements are desired. Small movements are required to help lessen the
chance of missing the optimum positioning.
Caution
Bulb can be
VERY HOT